Blood glucose-monitoring techniques originally developed to aid outpatient management of diabetic patients are now being used to facilitate hospital care. However, applications in hyperglycemic patients have been limited because many glucose-oxidase strips and meters respond only to glucose values ≤ 400 mg/dl. We asked if prior dilution of blood samples would permit reliable estimations. Ten consecutive decompensated diabetic patients (age 35–73, glucose 506–879, HCO3 12–28) had blood glucose determinations done simultaneously by the hospital laboratory and by Chemstrip bG after dilution of heparinized blood 1:2 in saline. Thirty-one samples were obtained before and during insulin therapy. Correlations with laboratory glucose values were 0.95 with strips read by Accu-Chek meter and 0.90 read visually, both P < 0.001. Average deviations from laboratory values were 7.9% with Accu-Chek and 12.9% with visual readings. Accu-Chek deviations averaged 9.6% for glucose > 700 mg/dl, and 6.9% for glucose > 400 mg/dl. Over the first hour of insulin therapy, glucose fell 150 ± 30 mg/dl by Accu-Chek, comparable to 168 ± 29 by laboratory measurement; the decrement by visual reading was 107 ± 32, not significantly different. We conclude that dilution of blood samples with glucose > 400 allows reliable estimation of elevated values by home glucose-monitoring techniques. This approach is cost-effective and provides the rapid feedback needed for the management of critically ill patients.